Quantus™ Fluorometer
The Quantus™ Fluorometer is a compact, dual-channel fluorometer designed to provide highly sensitive fluorescent detection when quantifying nucleic acids. It is optimized with preprogrammed settings for Promega QuantiFluor® Dye Systems.
Bio-Science Kft. is the official Hungarian distributor of Promega. We provide quotes, technical advice and service in Hungary - request a quote.
The Quantus™ Fluorometer is a compact, dual-channel fluorometer designed to provide highly sensitive fluorescent detection when quantifying nucleic acids. The Quantus™ Fluorometer is optimized with preprogrammed settings for Promega QuantiFluor® Dye Systems (QuantiFluor® dsDNA, RNA and ssDNA Systems) to quantitate nucleic acids and offers the flexibility to create customized methods and quantitation settings for other fluorescent dyes.
Features
- Easy, intuitive controls.
- Calculates concentration automatically.
- Store the last 50 readings and transfer data in real time.
- Recommended for NGS and compatible with Illumina (HiSeq/miSeq), Roche (454) and LifeTech (ion Torrent and ion Proton) Systems.
- Easily customize sample names, view concentration results, raw values, sample volumes and a quality check.
- Compact size at only 22.7 x 11.5 x 4.5cm.
- Large 3-inch liquid crystal display screen.
- Intuitive navigation.
Performance
Fluorescence-based nucleic acid quantification methods are more sensitive than absorbance-based methods, allowing you to quantitate low-level DNA samples accurately. Custom protocols can be created for the Quantus™ Fluorometer to detect even lower amounts.
Complement your absorbance-based methods and quantify your target molecule over a wider range of concentrations. Quantitation of more dilute samples is possible using more input dsDNA per assay.
Calibration and Sample Measurement
The calibration process eliminates the need to generate a standard curve for each set of unknown samples. Simply measure the fluorescence of your blank sample and the standard that represents the highest point on your standard curve. The Quantus™ Fluorometer saves the calibration data for future measurements.
No computer is necessary to use the Quantus™ Fluorometer, but after installing the free Quantus™ Software you can automatically transfer your quantitation results from the Quantus™ Fluorometer to a personal computer in real time.
Workflow
The high sensitivity makes the Quantus™ Fluorometer perfect when working with samples types such as formalin-fixed paraffin-embedded (FFPE) tissue that contain low levels of nucleic acid. You can confidently quantify nucleic acid from FFPE tissues and proceed to your downstream assays, knowing that there is DNA or RNA present.
Source: Promega product page (updated 2026-10-31).
| Parameter | Value |
|---|---|
| Size | 22.7 × 11.5 × 4.5cm (L × W × H) |
| Weight | 400 grams |
| Linear Dynamic Range | Five orders of magnitude, assay-dependent |
| Detectors | Solid-state silicon sensor |
| Excitation Filters | Red: 640nm shortpass (wavelengths up to 640nm); Blue: 495nm shortpass (wavelengths up to 495nm) |
| Emission Filters | Red: 660–720nm; Blue: 510–580nm |
| Calibration Type | Single-point (blank and standard) |
| Sample Tube | We recommend Axygen thin-walled PCR tubes (Axygen Cat.# PCR-05-C, available from Fisher or VWR) |
| Automatic Sleep Mode | After 1 hour of inactivity |
| Power Requirements | 5V, 0.2A Maximum |
| Operating System Requirements (for optional Quantus™ Software) | Windows® 10, Windows® 8, Windows® 7 (32- or 64-bit); .NET Framework 4.0 or later; At least one USB port available |
| PRODUCT | SIZE | CATALOG # |
| Quantus™ Fluorometer | 1 each | E6150 |
| PRODUCT | SIZE | CATALOG # |
| 0.5 mL PCR Tubes | 50 pack | E4941 |
| 0.5 mL PCR Tubes | 200 pack | E4942 |
| Quantus™ Instrument Standard Service Agreement | 1 each | SA4040 |