Dispensing Booth
Dispensing booth is also called sampling booth or weighing booth.
Reference picture
![]()
Reference picture
Air flow pattern
![]()
Differential pressure gauge
Dwyer Magnehelic gauges are used to measure the pressure drop across the HEPA filter
LCD Display
Microprocessor controller
![]()
PVC anti-static dustproof curtain.
![]()
Features
- Automatic variable-frequency system monitors the air velocity to ensure its stability in the work area.
- Differential pressure gauge is equipped to monitor the filters in real time.
- Uniquely designed air duct effectively controls the noise.
- Uniform flow design of the main working area protects the operators, prevents environmental pollution and cross-contamination of products.
- Intelligent control mode and alarm system ensure reliable operation.
- Smooth transitions between wall and floor eliminate blind spots.
- Customized designs are welcome.
|
Model |
BKDB-1200 |
BKDB-1800 |
BKDB-2400 |
|
|
Clean Level |
ISO 5 (Class 100), Class A |
|||
|
HEPA Filters |
99.999% efficiency at 0.3 um |
|||
|
Air volume |
Supply air volume: ≤7500m3/h, Exhaust air volume: ≤2250m3/h; Adjustable
|
|||
|
Material |
Fully stainless steel |
|||
|
Air Velocity |
0.35~0.65m/s, adjustable |
|||
|
Noise |
≤75db |
|||
|
Illumination |
≥800lux |
|||
|
UV Lamp |
Emission of 253.7nm |
|||
|
Power Supply |
110/220V±10%, 50/60Hz |
|||
|
Consumption |
400W |
800W |
800W |
|
|
Weight |
150kg |
200kg |
250kg |
|
|
Work Area Size (W*D*H) |
1200*1200*200mm |
1800*1800*2000mm |
2400*1800*2000mm |
|
|
Customized design for clean level, size or material is available |
||||