Temperature Uniformity – the growth chamber is surrounded by heated water and monitored using dual temperature sensor probes. Model NU-8600 offers temperature uniformity within 0.20°C at 37°C.
CO2 Gas Accuracy - using a dual wave infrared (IR) sensor that is insensitive to other components, such as water vapor, the chamber maintains CO2 levels accurate within 0.1%.
Growth Condition Recovery – Quicker and more stable, In-VitroCell CO2 Incubators recover gas and temperature back to set point faster than other CO2 Incubators.
NuTouch Electronic Control System (ECS) – easily control system parameters with the touch of a finger. The NuTouch ECS is a user-friendly touchscreen LCD that allows for the control of
parameters and offers status indicators, on screen instructions, and notifications to assist with proper use.
Hypoxic Conditions – model NU-8631 provides the ability to suppress oxygen in the growth chamber by injecting N2 gas to meet set point by monitoring and controlling oxygen using a
sensor.
Features
Constant Contamination Control
Closed Loop HEPA Filtration – Maintained at positive pressure, gas and air entering the growth chamber continuously pass through 99.99% @ 0.3 micron efficient HEPA filters producing an
environment similar to an ISO Class 5 clean room, slowing airflow to one chamber volume air change per 30 minutes to minimize cell desiccation.
Coved interior corners – A crevice-free interior is easier to clean and eliminates potential growth areas within the chamber.
CuVerro Antimicrobial Surfaces – add CuVerro® Antimicrobial Copper Shelving (optional) - to the incubator growth chamber and shelving to kill bacteria* to minimize potential contamination. CuVerro® is laboratory tested and EPA registered. CuVerro® Antimicrobial Copper Surfaces kill more than 99.9% of bacteria* within two hours, and continue to kill 99% of bacteria* even after repeated contamination and regular cleaning.
In-VitroCell ES CO2 incubators provide stable, contamination-controlled conditions for cell and tissue culture.

Standard Features
NuTouch Electronic Control System
Closed Loop HEPA Filtration System
100% Stainless Steel Coved Interior Chamber
Dual Temperature Sensor Probes
Infrared (IR) CO2 Sensor
O2 Control System (NU-8631)
Four (4) Stainless Steel Shelves
Eight (8) Stainless Steel
Shelf Guides
Four (4) Wall Brackets
Heated External Right Hinged Door Swing (Field Reversible)
Inner Right Hinged Door Swing (Field Reversible)
Remote Alarm Output Contacts
RJ-45 4 to 20 mA Analog Output
RJ-11 Communication
RS-485 Communication
USB Port
CO2 Sample Port
Water Fill Port
Water Drain Valve
Adjustable Leg Levelers
Access Port
One (1) Stainless Steel Water Pan
One (1) 8 ft. / 2.5 m
Electrical Cord
Temperature Control System
Temperature Sensor Type:
Precision Integrated Circuit
Default Set Point: 37°C
Chamber Temperature Range:
5°C to 55°C (5°C Above
Ambient to 30°C Max. Ambient)
Chamber Temperature
Uniformity: ± 0.20°C @ 37°C
Temperature Accuracy: ± 0.1°C
Temperature Recovery:
0.12°C/Minute Average
Temperature Display Resolution:
0.1°C
Minimum Qualifications for
Sterilization:
145 DEG Cycle 135°C
95 DEG Cycle 85°C
Electrical Requirements
Startup Power: 625 watts
Running Power: 250 watts, 60 Hz
Heat Rejected: 14 BTU / min.
Utility Connections
Gas Connections: 0.25 in.
(6.3 mm) Tubing Connections
Gas Input Pressure: 20 PSIG
(1.4 BAR) Input Pressures
Maximum. Two-Stage Gas
Regulators Required.
CO2 Control Systems
CO2 Sensor Type:
Infrared Single Source Dual
Wavelength
CO2 Control Logic:
Fixed Algorithm / Manual
Environmental Adaptable.
Default Set Point: 5%
CO2 Range: 0.1 to 20%.
(0.0 Set Point Idles System)
CO2 Accuracy: ± 0.1%
CO2 Recovery: Up to 5%
-0.50% / +0.20% in
5 Minutes Average.
CO2 Display Resolution: 0.1%
O2 (8631)
Zirconia Ceramic Sensor
Default Set-Point: 21%
O2 Range: 0.5 to 21%
O2 Accuracy: ± 0.25%
O2 Recovery: 5% ± 0.5% / 20 min.
Optional Features
Incubator Stacking Brackets
Coiling Coils
Additional Stainless Steel
Shelves with Guide Brackets
CuVerro® Antimicrobial Copper
Shelving and Guide Brackets
CuVerro® Antimicrobial Copper
Water Pan
Automatic CO2 Tank Switch
(External)
Left Hinged Door Swing
Gas Tight Sectioned Inner Door
Platform with Castors
Moisture Proof Duplex
CO2 Regulator (2 Stage)
N2 Regulator (2 Stage)
CO2 Analyzer Fyrite Kit
(Dry) 0-20%
Replacement Fluid for
CO2 Analyzer